Influence of the fiber optic length on the repeatability of the Near Infrared Spectroscopy signal from animal protein by-products
The available NIRS scientific literature concerning the analysis of animal by- product proteins meals (ABPs) has shown that it is possible to chemically characterize these interesting protein sources for animal feed, and also to detect and quantify the species composition in a particular type of meal. Since, all work carried out so far refers to the analysis performed “at line”, it is still necessary to continue the scientific work that can show to - the producing industry (rendering plants), to its customers (feed manufacturers) and especially to whose responsible of the official inspections of ABPs, that is feasible to analyze -in real time- large volumes of sample, using on- line instrumentation. Traditionally, fiber probes used for NIRS analysis of agri-food products are between 1 to 3 meters length. More recently, instruments coupled to fiberoptic probes up to 100 meters length have appeared on the market, but not much information about their potential applicability has been reported. The light power propagating in a fiber decays exponentially with the length due to absorption and scattering losses, and therefore, the amount of light reaching to the detectors is also depending of the properties of the samples analysed. The aim of this work is to assess the effect of the fiber optic probe length on the noise and repeatability of the NIRS spectral signals from ABPs.
A total of 10 samples representatives of the ABPs from several species where analysed in a MATRIX-F FT-NIR instrument, interfaced to a detection head connected by a fiber optic cable of different lengths (5, 50 and 100 m). Ten spectra were collected for each sample and each fiber length. The noise level along the full spectral range available from the spectrometer (834.2–2502.6 nm) was evaluated pre-treating the signal using first derivative. Spectral repeatability was evaluated using the Root Mean Squared (RMS) statistic, being defined as the averaged root mean square of differences between the subsamples scanned at n wavelengths. For each fiber length and sample, the RMS for an individual subsample (j) of the sample (k), and the MEAN values for a given k sample were calculated using Excel.
Results reveal that noisy regions appear at the beginning and at the end of the spectral region, being the optimum working ranges: 1180-2370, 1180-2200 and 1180-2100 nm for probes of 5, 50 and 100 m., respectively. Furthermore, the spectrum becomes more noisy as the fiber length increases and at higher wavelengths 2100-2370 nm. The spectral repeatability results showed that once the regions of measurement were optimized for each fiber optic length, the values for the statistic RMS (MEAN) were very similar, being 15.600, 15.055, 13.417 μlog1/R for 5, 50 and 100 m fiber lengths, respectively.
The results obtained show the feasability for performing the on-line analysis of ABPs, placing the FT- NIRS instrument up to 100 m distance of the measuring point, without appreciable loss of spectral repeatability. Further work is in progress to develop calibrations for the chemical and species characterization of ABPS using this analysis mode.