Matching portable NIRS instruments for on farm monitoring of the protein and fat contents of cow milk.
In the near future more and more dairy farms will uptake more or less sophisticated Precision Livestock Farming (PLF). In the case of a dairy farm automated feeding systems, milking robots and stable and farm management systems (i.e. cameras, microphones….) which will bring up to the farmer greater and higher productivity, increasing animal welfare and making easier farm operations. However, Precision Dairy Feeding (PDF) is not only a matter of automation devices and computers programs, PDF is changing the way we manage cows by allowing us to monitor the animal more closely for improved nutrition and health, performance, reproduction, and well-being. Affordable methods to analyze the ration components and also the quality of the milk of every cow are essentials as PLF management tools. Near Infrared Spectroscopy has demonstrated its ability to analyse both feed and milk using instruments located in a laboratory. Nutrition and feeding management practices of individual dairy cows offer a means of making rapid changes in milk composition, but the relationship between feed constituents and milk composition is complex. The greatest changes can be brought about in the concentration of milk fat. Far smaller changes are possible in milk protein concentration and very small changes in lactose concentration occur sometimes in response to diet. Whatever the technical developments, economic factors will be the final determinant of the uptake of such techniques by the dairy farmer. The instantaneous knowledge of individual dairy milk composition changes can be used both as an indicator of animal health and metabolic changes, and as additional information for dairy producers.
The main goal of the present work is to demonstrate the ability of hand-held NIR spectrophotometer for quality control of milk analysis and further the transferability of developed models along portable NIR instruments.
A total of 128 fresh milk samples were collected from different cows of the experimental farm located in SERIDA between 2014-2015 and scanned using the hand-held MicroPHAZIR TM from Thermo Scientific working in a wavelength range of 1600-2400 nm to develop the calibration model (Master Unit). Additionally, 48 milk samples were scanned in two different units of MicroPHAZIR from SERIDA and UCO. Ten samples were selected to test standardization procedure and the rest (N=38) to validate the transference process.
The cross-validation statistics of calibration models displayed great predictive ability for fat and protein content with SECV of 0.087 and 0.115 % and r 2cv of 0.959 and 0.750 respectively. When the equations were applied to a second unit there are a great loss of performance with SEP of 1.102 and 0.229 % and SED of 1.031 and 0.276 for fat and protein, respectively. After applying the developed standardization matrix, the SED between SERIDA and UCO reached a great reduction of 85 % and 40 % for fat, and protein respectively. These SED values are around 1.5 times of SECV.
It can be concluded the ability of portable NIR instruments to estimate the individual dairy milk composition changes and the promising results of transference procedure between these hand-held instruments.