Investigating inverse and direct 13C detect experiments and slow spinning of in vivo multiphase samples
Rajshree Ghosh Biswas
Department of Physical and Environmental Science / Scarborough Campus / University of Toronto
Now you could share with me your questions, observations and congratulations
Create a topicWith the ever-increasing presence of pollutants in the environment, a non-invasive analytical approach is required to understand their toxic impacts on organisms and their ecosystems. Comprehensive Multiphase (CMP) NMR spectroscopy combines hardware from both HR-MAS and solid-state NMR into one probe, allowing for real-time observation all phases (solids, liquids, and gels) in unaltered samples.1
The CMP E-Free MAS I-Probe is a two-coil design with an inner solenoid coil dedicated to 13C and an outer cage coil for decoupling. This can provide over 100% increase in 13C SNR relative to a multiple tunned single solenoid CMP coil. The outer 1H coil reduces RF heating and is ideal for multiphase analysis of in vivo samples which are suspectable to RF heating. In toxicity research, living organisms act as biosensors, with their metabolic profile (HR-MAS) providing vital, real-time information regarding the toxic-mode-of-action of the contaminants, whereas structural changes (shell thinning in crustaceans, i.e. solids-state), can be indicative of longer term non-lethal exposure.
In 2D NMR-based metabolomics, 13C is often observed in the indirect dimension due to its lower sensitivity relative to 1H. However, during in vivo NMR, a major drawback of this is the detection of a large water resonance (and its spinning side bands at slow MAS, required to keep the organisms alive) in the direct dimension, masking important metabolites. As carbon makes up the backbone of almost all natural compounds, 13C NMR can be advantageous for observing intact complex environmental samples. With the increased 13C sensitivity of the I-Probe, direct detection of 13C is more feasible, allowing for a new suite of 13C detect experiments to overcome the limitations of the traditional 2D 1H detected experiments. This present work compares inverse and direct 13C detect 2D NMR experiments of Hyallela azteca (freshwater shrimp, commonly used in toxicity testing) to determine which is best for in vivo NMR.
Fig.1B and D depict a large water signal in the HSQC, masking the carbohydrate resonances. This becomes worse at slower speeds, where spinning sidebands, arising from the water, dominate, making metabolite assignment almost impossible. Though faster spinning can help to move the sidebands out of the spectral window, survival time decreases drastically. However, in the HETCORs (1A and 1C) where 13C is directly observed, no water signal is detected, thus providing clean spectra for assignments. Fig.2 depicts the first report of an in vivo HETCOR of a single 13C-enriched H. azteca at 2500 Hz. The major metabolites are assigned and can be used to observe biochemical changes in response to environmental stressors. Thus, with improved 13C sensitivity of the new CMP I-Probe, quality 2D 13C NMR can be obtained, providing complementary information to 1H observed in vivo NMR.
Lorena Mara Alexandre e Silva
Hello Rajshree, congratulations on your work and presentation. Is this like a HRMAS probe with CMP technology? What is the volume of the rotor you used? Do you flow water into the rotor to keep the organisms alive?
Elenilson de Godoy Alves Filho
Dear Rajshree, congratulations on your great work! It is a very interesting and promising study!
Please, could you tell me if it is possible to acquire spectra under quantitative parameters using this probe?
Best Regards!
Rajshree Ghosh Biswas
Hi Elenilson,
Thank you!
This probe can do quantitative measurements so as long the parameters are set.
Thank you for your question!
With nearly 200,000 papers published, Galoá empowers scholars to share and discover cutting-edge research through our streamlined and accessible academic publishing platform.
Learn more about our products:
This proceedings is identified by a DOI , for use in citations or bibliographic references. Attention: this is not a DOI for the paper and as such cannot be used in Lattes to identify a particular work.
Check the link "How to cite" in the paper's page, to see how to properly cite the paper
Rajshree Ghosh Biswas
Hi Lorena,
Thank you!
CMP probes tend to be solid state probes with HR-MAS hardware.
I used a 80ul rotor for the studies in the presentation. For the in vivo samples, we drill a hole in the rotor cap to allow air exchange, but the entire rotor is filled with water to keep the organism alive.
Thank you for your questions!
Lorena Mara Alexandre e Silva
Hi Rajshree thanks for your feedback. Cheers