Testes de aquecimento de equipamentos utilizados para medições radiométricas em laboratório
The first tests of the Goniometer by LaRaC (Laboratório de Radiometria e Caracterização de Sensores Eletroópticos) are presented, with focus on its equipments functionality. First results showed anomalous behavior of the reflectance factor spectrum of a reference panel obtained at the end of a BDRF data colleting cycle. Since this, exploratory measurements were conduced to identify possible changes in the Spetroradiometer and light source behaviors during its warm up period. To investigate more precisely the influences of each equipment, individuals warm up tests were made, whose results indicated higher variations in the radiance spectral data due to changes in spectroradiometer functionality, meanly in wavelengths regions associated with the detector overlap regions (i.e. 1000 nm and 1800 nm). It was observed that the spectroradiometer achieves reasonable stability after a warm up period of 2,0 h, while the light source showed low variability. In order to verify possible correlations between data variability and environmental parameters, a thermo hygrometer was used to collect data related to these parameters, such as temperature and gases concentration, at the same time as the tests were made, whose results showed no correlation between them. Finally, data in a wavelength band was selected from individual tests to compare radiance in band data to spectral radiance data and, as result, the same behavior was observed in both data.