INVESTIGATION OF SPECTRAL INTERFERENCES IN THE DETERMINATION OF SELENIUM IN COCONUT WATER BY ATOMIC ABSORPTION SPECTROMETRY

Favorite this paper

Interferences in graphite furnace atomic absorption spectrometry (GF AAS) are classified as non-spectral or spectral. The spectral interferences result from direct atomic line overlapping, scattering of the radiation on solid or liquid particles, and radiation absorption by gaseous molecules1,2. Molecular absorption results in (i) electron excitation spectra, which are electron transitions between electronic levels and exhibit a characteristic fine structure; or (ii) broad bands that are continuous signals due dissociation of the molecule into atoms, designated as dissociation continua1. The goal of this study was the investigation of spectral interferences in the direct determination of Se in coconut water by atomic absorption spectrometry. Measurements were carried out in an AAS 5 EA model GF AAS spectrometer (Analytik Jena) equipped with a D2 lamp as background correction (BC) and a Se hollow cathode lamp (IST) at wavelength 196.0 nm. A high-resolution continuum source AAS spectrometer model contrAA 700 (Analytik Jena), equipped with transversely heated graphite tube atomizer (HR-CS GF AAS) was employed to identify the spectral interferences. It was observed a peak-like background during the direct analysis of coconut water samples by GF AAS with D2 as BC system, which resulted in overcorrection of the signal. In order to identify the interference, the same sample was analyzed using HR CS GFAAS at 196.027 nm. The same background profile was identified as a diatomic molecule, which overlapped the analytical signal of Se. Several molecules were investigated, based on the sample composition, and it was identified as NO molecule (generated using 100 µg of NH4NO3). The NO is probably generated from proteins during the atomization step. HR CS GF AAS allowed the identification of the spectral interference observed by GF AAS in Se determination in coconut water, which could be overcome by sample preparation, such as dilution or digestion. However, it implies worse sensitivity due to sample dilution during procedure. Chemical vapor generation techniques convert the analytes to volatile species, separating them from the matrix and also allow preconcentration, enhancing the sensitivity. Therefore, this technique could be an alternative to overcome matrix interferences.