DETERMINATION OF SILICON AND ALUMINUM BY WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETRY USING WHITE CLAY AS STANDARD CALIBRATION

Favorite this paper

Trace amounts of Si and Al in crude oils may affect the quality of the refined products. The goal of this work is to develop a method to quantify lower amounts of these elements in crude oil by wavelength dispersive X-ray fluorescence spectrometry (WDXRF) using a proper calibration strategy. A reliable determination of Si and Al in crude oil could not be made properly using analytical curves made with analyte organic standards as these elements are mostly present as aluminosilicate particulates dispersed in the sample. Crude oil were determined directly without any pretreatment, using 100 µL of sample spiked in the center of a filter paper that was then placed into the instrument cell for measurements. Sample homogenization procedure was studied in order to enable accurate results. Quantification was made by analytical curves in the concentration range up to 30 mg kg-1 for Si and up to 10 mg kg-1 for Al using standards made by dispersing a reference white clay in mineral oil/tetrahydrofuran. White clay allowed a proper correlation of results since it simulates the form how Al and Si can be found in the oil. The quantification limits (10 times the standard deviation, n = 10) for Si and Al in the oil were respectively 0.5 and 0.2 mg kg -1. Matrix interferences were minimized once the sample formed a thin film onto the substrate that was dried prior to analysis. Tests using crude oil fortified with clay enable recovery results above 90%. For real samples, the results for the elements (present as particles in the size range 200 to 6000 angstrons as determined by X-ray diffraction analysis) were statistically comparable to the ones obtained by flame atomic absorption spectrometry despite the fact that these concentrations were in the lower range of the analytical curve, where uncertainty tends to be higher.