55155

THE FEASIBILITY OF DETERMINATION OF Pb IN VOLATILE FUELS BY HIGH-RESOLUTION CONTINUUM SOURCE GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY USING DRIED MATRIX SPOT SAMPLING

Favorite this paper

In this work a simple and practical method for the determination of Pb in volatile fuels by solid sampling high-resolution continuum source graphite furnace atomic absorption spectrometry was investigated. In contrast to the conventional methods, which required several sample preparation’s steps, the proposed method uses pieces of filter papers to preconcentrate and store Pb and others trace elements from volatile fuel samples. This proposal is based on the precut–dried matrix spot sampling, in which a volume of sample is deposited and dried onto a filter paper disk previously adapted into a PTFE cell, and then it was cut in small pieces for trace elements determination.