Determination of the antioxidant capacity of Primulae flos cum calycibus with NIR and ATR-IR spectroscopy including an evaluation of reference imprecision impact on the SEP

Favorite this paper

In phytopharmaceutical industry a high quality of the processed plant material is crucial to obtain a high quality end product. Primulae flos cum calicybus is a traditional medicinal plant used in industry and antioxidative assays are good indicators to determine its quality. The disadvantage of wet chemistry assays is their time consuming extraction and measurement steps as well as their poor precision. Therefore in this work fast near-infrared (NIR) and attenuated-total-reflectance infrared (ATR-IR) spectroscopic methods in hyphenation with partial least squares (PLS) regression were deployed to determine the antioxidant capacity of Primulae flos cum calycibus samples. Five standard assays were used as reference methods for the measurement of the antioxidative capacity. As these assays exhibit poor precision results, the influence of reference imprecision on the SEP and RPD was evaluated. For this, in a first step the SEPs and the RPDs were corrected for the imprecision of the reference methods to obtain SEPcorr and RPDcorr values. In a second step a reference based error (RBE) was calculated which expresses the percentage of the SEP caused by reference imprecision. With help of the RBE it can be monitored if the reference is precise enough to obtain reliable error estimations in a given application.
A Buechi NIR-Flex N-500 FT-NIR (Buechi, Flawil, Switzerland) together with NIR Ware 1.4.3010 software (Büchi, Flawil, Switzerland) and a Perkin Elmer Spectrum 100 (Perkin Elmer, Waltham, USA) in combination with a Spectrum software version 6.3.1.0134 (Perkin Elmer, Waltham, USA) were used to collect the spectra. 32 Primulae flos cum calycibus samples were milled with a Retsch ZM 200 mill (Retsch, Haan, Germany) to a particle size less than 0.5 mm. The samples subsequently were measured with spectroscopy and Folin-Ciocalteu, FRAP, DPPH, ABTS and CUPRAC assays. NIR spectra were pretreated with SNV, 1st or 2nd derivative Savitzky-Golay and combinations of SNV and a derivative. ATR-IR spectra were normalized between zero and one and afterwards pretreated with 1st or 2nd derivative. The best pretreatment was determined by high R2 and low SEP values.
Both spectroscopic methods exhibited best predictive power for the ABTS assay and NIR spectroscopy demonstrated in general advantages over ATR-IR in this study. For the test-set validation of the ABTS assay the NIR reached a R2 (RPDcorr) of 0.94 (4.66) and the ATR-IR resulted in a R2 (RPDcorr) of 0.94 (4.10). The RBE values of 8.56% for NIR and 7.30% for ATR-IR were both in an acceptable low range. This indicates that the reference was precise enough for this application. Supported by this data, the feasibility of vibrational spectroscopy as a fast and simple tool to replace wet chemistry assays for the measurement of the antioxidant capacity of Primulae flos cum calycibus samples was demonstrated. The developed methods are considered to be useful improvements for phytopharmaceutical industry.