The Use of Near-infrared Spectroscopy and Chemometrics to Monitor the Drying of Heparin

Favorite this paper

Near-infrared (NIR) spectroscopy is a powerful tool which is broadly utilized for the understanding and monitoring of processes. In this study, NIR spectroscopy combined with chemometrics was adopted to demonstrate the feasibility of qualitative and quantitative analyzing the heparin drying process. The aim of this study was to assess whether NIR spectroscopy is capable of: (i) qualitative discriminating abnormal samples from normal samples for the fault diagnosis; (ii) quantitative determining the loss on drying (LOD) values of heparin during the drying process.
In this work, twelve normal aliquots of heparin with LOD value lower than 5% and three abnormal aliquot with LOD value higher than 5% were obtained to evaluate the ability of fault diagnosis. The spectra of heparin samples collected during the drying process were acquired using an Antaris II Fourier-transform near-infrared spectrophotometer (Thermo Fisher Scientific, USA) with an integrating sphere attachment. The wavenumber range and resolution are 10000–4000 cm-1 and 4 cm-1, respectively.
Processing and analysis of spectra were carried out with Matlab 2010a (Version 7.0.10 Mathworks, Natick, MA, USA) and PLS toolbox (Eigenvector Research, Inc., Manson WA, USA). The qualitative and quantitative model information was listed below.
1. Qualitative analysis of data
Principal component analysis (PCA) was adopted for data dimension reduction to develop the MSPC charts. First, spectra were pre-processed with SNV combined with the mean center to remove useless information. Then, eight normal aliquots were selected as the calibration set to build the qualitative model; four normal aliquots and three abnormal aliquots were taken as the validation set to evaluate the ability of fault diagnosis. The scores of PC1, Hotelling’s T2, and the Q residual control charts of normal aliquots were constructed to monitor the drying process. The results showed that PC1 and the Q residual control charts could accurately distinguish abnormal from normal aliquots. Hotelling’s T2 control charts couldn’t detect the abnormal conditions.
2. Quantitative analysis of data
Quantitative model was established by the PLS algorithm. First, the second-derivative with Savitsky-Golay 15 points smoothing combined with autoscaling was applied to the spectra. Then, 112 samples from 8 batches were selected as the calibration set, and 56 samples from 4 batches were used as validation set to evaluate the predictive ability of the model. Next, three different variable selection methods, including the correlation coefficient, interval partial least squares (iPLS) and successive projections algorithm (SPA), were respectively used to select characteristic variables to improve the predictive ability of the model. The results showed that the optimal model was obtained by the regions of 9011.7-8628.0 cm-1, 8240.3-7856.6 cm-1, 7469-6699.5 cm-1, 5540.5-5156.7 cm-1 and 4383.4-4000 cm-1, which were selected by Forward-iPLS. The values of R2, RMSEC, RMSECV, RMSEP and RPD of model are 0.976, 0.419%, 0.648%, 0.902% and 4.68, indicating that the model was able to accurately determine the LOD (%) of heparin during the drying process.