34140

USE OF INFRARED MICROSCOPY IN THE DETECTION OF PARTICLES OF ANIMAL ORIGIN IN BONE ASHES.

Favorite this paper

Bovine Spongiform Encephalopathy (BSE) is a bovine mortal disease. Epidemiologic studies linked the disease to contaminated animal feed with infected protein (prion).
In Argentine is banned to feed bovine with different components of animal origin, except bone ashes and fish meal, among others. This product is obtained from animal bone tissue calcination and it should not contain organic matter.
Nowadays, optical microscopy is used to analyze animal feed and its components. In this method, technical analyst expertise is very important. This is why the development of a methodology that shows results that can be rapidly and automatically compared is still a pending issue.
The specific infrared microscopy method presented in this work delivers qualitative results in terms of the presence or absence of animal particles by applying specific decision rules.
The object of this work is to assess, by means of several tests, the methodology of NIRM for the qualitative determination of the presence of particles of animal origin in matrices of bone ashes.

The spectra were collected by a Thermo Scientific Image Microscope coupled to Fourier Transform Infrared Spectrometer in the reflection mode. A liquid nitrogen cooled detector was used. Aperture size was 100 × 100 microns and spectral range was 6000cm-1 - 3000cm-1.
Ten batches of bone ash samples were prepared and fortified with the following levels of bone fragments: 0.5%, 0.1% y 0.05%, 0.02% y 0.01%. Bone fragment was used as analyte because is the most representative meat and bone meal component

In addition, matrices of pure ashes were analyzed to rule out false positives. For this purpose, 20 ± 0,01 g of each sample were poured on fitted sieves of 125μ and of 300μ, ordered in a decreasing size according to the mesh. It was carefully shaked to distribute the particles according to the size. Then, the sample was quartered twice.

Four mappings were done per sample, obtaining 1000 spectra per mapping from 1000 dots on the slide. They were selected from an area on the slide of 50 dots in the axis X and 20 dots in the axis Y. Spectra were automatically collected at these points.

In order to find the fragments of animal origin in the sample, specific software was developed. The program analyses the spectra and evaluates if they fulfil the published criteria. The graphical interface was developed in Delphi 4. The module of logic was developed in php.

From the validation results it can be inferred that 4000 particles should be observed in order to achieve a detection limit of 0,05 g bone fragments/100 g bone ash. Selectivity, intermediate precision and repeatability were also analyzed with satisfactory results. Therefore, the validated method can be easily applied in routine analysis of bone ashes.