Is NMR Sensitive Enough to Analyze Unconcentrated Wastewater? Direct, Real-time Monitoring of Ozone and UV Treated Wastewater Using Flow NMR

Favoritar este trabalho
Como citar esse trabalho?
Detalhes
  • Tipo de apresentação: Poster Presentation
  • Eixo temático:
  • Palavras chaves: wastewater treatment; Organic composition; Unconcentrated wastewater;
  • 1 Department of Physical and Environmental Sciences / Environmental NMR Centre / University of Toronto
  • 2 Environmental NMR Centre / University of Toronto
  • 3 Synex Medical
  • 4 Chemistry / Scarborough Campus / University of Toronto
  • 5 Department of Physical and Environmental Science / Scarborough Campus / University of Toronto
  • 6 Chemistry / Environmental NMR Centre / University of Toronto
  • 7 Environmental Monitoring & Reporting Branch / Ontario Ministry of the Environment
  • 8 Bruker
  • 9 Department of Physical and Environmental Sciences / Scarborough Campus / University of Toronto
  • 10 University of Toronto

Is NMR Sensitive Enough to Analyze Unconcentrated Wastewater? Direct, Real-time Monitoring of Ozone and UV Treated Wastewater Using Flow NMR

Katelyn Downey

Department of Physical and Environmental Sciences / Environmental NMR Centre / University of Toronto

Resumo

Globally, clean freshwater is becoming increasingly scarce, and effectively treating wastewater for reuse is therefore a critical feat, with new disinfection methods (ozonation, photo-irradiation, etc.) abounding. Although NMR spectroscopy can provide an untargeted overview of wastewater’s chemical composition in real time, its low sensitivity generally precludes it from analysis of unconcentrated samples.
The objective of this research was to evaluate whether NMR spectroscopy is sensitive enough to analyze unconcentrated wastewater and monitor removal of the organic load. A second aim was to use online NMR spectroscopy to evaluate three different treatments (ozonation, UV irradiation, ozonation + UV irradiation) on samples of post-primary, post-secondary, and final effluent wastewater fractions. A closed-loop flow system was used to monitor chemical classes and individual compounds over a 51-hour period while treating a sample of one of the wastewater fractions with UV irradiation from a sunlight simulator and/or O3 from an ozone generator (Fig. 1). The sample flowed from the treatment tube into a 500 MHz NMR and 1D 1H NMR spectra were recorded [1].
Experiments aimed to assess the efficiency of organic load removal by each of the treatments on the three wastewater fractions. In general, ozone + UV irradiation degraded organics to the greatest extent, and UV irradiation alone was least efficient. One exception was that ozone alone most effectively removed organics from post-primary wastewater (Fig. 2), potentially because in the ozonation + UV irradiation treatment, the light also degraded the ozone. Alternatively, post-secondary wastewater and final effluent contain more recalcitrant compounds (i.e. difficult to degrade by ozone alone), and the combination of ozone and UV irradiation was most effective likely because the UV light produces reactive OH radicals from the ozone. Results also showed that specific compounds undergoing a single treatment may be removed at differing efficiencies depending on the wastewater fraction analyzed, suggesting that degradation is partially influenced by the wastewater fraction.
In future wastewater research, it may be beneficial to assess the effectiveness of a treatment wherein wastewater is initially exposed only to O3, and subsequently irradiated to produce OH radicals, potentially optimizing treatment efficiency. Additionally, smaller and more affordable low-field NMR spectrometers hold great promise for future on-site wastewater analysis at treatment plants for real-time monitoring.

Questões (2 tópicos)

Compartilhe suas ideias ou dúvidas com os autores!

Sabia que o maior estímulo no desenvolvimento científico e cultural é a curiosidade? Deixe seus questionamentos ou sugestões para o autor!

Faça login para interagir

Tem uma dúvida ou sugestão? Compartilhe seu feedback com os autores!

Autor

Katelyn Downey

Hello Lorena, 

Thank you for your interest! We used a water suppression pulse sequence called Shaped PResaturation W5 Water suppression by GrAdient Tailored Excitation (SPR-W5-WATERGATE).

Autor

Katelyn Downey

Hello Elenilson, 

Thank you! Yes, the 51hr duration was used not to reflect a real wastewater treatment time period, but was simply chosen because it was long enough that we could track the kinetics of organic matter degradation (and follow them to completion).