Eletrodetecção da proteína C reativa: um marcador de risco cardiovascular
Cardiovascular diseases are the main cause of death around the world [1]. Creactive protein (CRP) is an important marker of cardiovascular risk due to its association and contribution to the atherosclerotic event [2]. The rapid diagnosis of cardiovascular diseases is important to prevent further complications. Thus, biosensors are potential alternatives, mainly due to their low cost, precision, specificity and portability. The aim of this research is the development of a bioelectrode for the detection of CRP in serum of patients with atherosclerotic disease. Therefore, graphite electrodes were functionalized with poly (3-hydroxyphenylacetic acid) and sensibilized with anti-CRP. Assay using positive and negative serum for atherosclerotic disease were carried out using differential pulse voltammetry (DPV) and atomic force microscopy (AFM). It was observed (Figure 1a) a decrease in the current signal amplitude about 1.5 times for EG/poly (3-HFA)/anti-PCR in relation by positive serum sample. Since C-reactive protein presents an isoelectric point of 5.45, in pH7.4 (assay condition) the charge net of the protein is negative, resulting in a decrease in electron transference, in K3Fe(CN)6/K4Fe(CN)6∙3H2O redox pair. The presence of the negative serum does not causes modification in the current response. The surface morphological analysis (Figure 1b) is in concordance with the electrochemical results.