Use of digital images and smartphone to determine soluble proteins in Whey protein using the Bradford method

Favorite this paper
How to cite this paper?
Details
  • Presentation type: Exposição de Pôster
  • Track: Química Analítica - ANA
  • Keywords: PhotoMetrix; colorimetry; Total soluble proteins; Sports supplements; digital image;
  • 1 UFSCar
  • 2 Universidade Federal de São Carlos

Use of digital images and smartphone to determine soluble proteins in Whey protein using the Bradford method

Bruna Soares Dionizio

UFSCar

Abstract

Currently, there has been an increase in the consumption of sports supplements such as whey protein, consisting of proteins derived from milk, by practitioners of physical activity, with the aim of increasing the synthesis of muscle proteins1. One of the colorimetric methods used to determine total proteins is the Bradford method using UV-Vis spectrophotometer2. The use of digital devices such as cameras and smartphones and PhotoMetrix®  software has been employed to perform colorimetric detection, aiming at the speed, practicality and low cost of analysis when compared to the spectrophotometer3. Thus, a study was developed in order to determine the total soluble proteins in 9 samples of sports supplements using digital images and PhotoMetrix®. To prepare the sample, 100 mg of each sample was solubilized in 10 mL of ultra-pure water and centrifuged. The calibration curve was prepared with 0 to 0.8 mg mL-1 of bovine serum albumin (BSA) protein. An aliquot of 20 µL of each sample was added to 1000 µL of Bradford's solution and homogenized. After 5 minutes of incubation (25 °C), the absorbance can be read at 595 nm using a spectrophotometer (reference method), as well as digital images were acquired with a 13 MP camera from a Samsung J500-M cell phone, then processed by the PhotoMetrix®. Addition-recovery experiments were performed with 3 samples, adding 1 mg mL-1 of BSA, to assess the accuracy of the method. To analyze the protein profile, supplement samples were submitted to gel electrophoresis. The analytical performance parameters obtained with the reference method were limits of detection (LOD) and quantification (LOQ) of 0.003 mg mL-1 and 0.01 mg mL-1, respectively of total soluble proteins, and RSD values in the range of 1-11%. When using PhotoMetrix® for protein quantification, the LOD and LOQ obtained were 0.0002 mg mL-1 and 0.0007 mg mL-1, respectively. The method demonstrated good precision of the measurements with RSD in the range of 1-13%. The addition and recovery experiments demonstrated good accuracy for both methods, with recoveries of 82-112% using PhotoMetrix®, and 84-106% using the reference method. It was observed that the LOD and LOQ values determined by PhotoMetrix® were 15-fold lower than the values obtained by the reference method. The concentrations of total soluble proteins (n = 3), determined in the 9 samples of commercial sports supplements ranged from 17±1 to 253±12 g kg-1 using the reference method, while PhotoMetrix® presented results between 14±1 to 304±33 g kg-1, demonstrating that the values determined by both strategies are in concordance. Besides that the formulations present great variability of concentration of total proteins solutions. Through the electrophoretic gel it was possible to observe that the sample #1 (S1) differs from the other samples qualitatively, and this result corroborates with the determined concentrations: this samples presented the lowest protein in its formulation. Thus the proposed procedure using PhotoMetrix® proved to be an excellent analytical tool for the determination of low concentrations of soluble proteins using the Bradford method.

[1] F.K. Haraguchi et al., Rev. Nutr., 19(4) (2006) 479-488.

[2] M.M. Bradford, Anal. Biochem., 72 (1976) 248.

[3] PhotoMetrix Database: http://www.photometrix.com.br.

Questions (2 topics)

Share your ideas or questions with the authors!

Did you know that the greatest stimulus in scientific and cultural development is curiosity? Leave your questions or suggestions to the author!

Sign in to interact

Have a question or suggestion? Share your feedback with the authors!

Author

Bruna Soares Dionizio

Olá Célia!

Muito obrigada,

Acreditamos que a sensibilidade está relacionada ao fato que utilizando o Photometrix tem-se apenas o processamento da imagem digital adquirida, ou seja, está relacionada com a aparência da solução. Enquanto que no espectrofotômetro há interação da radiação eletromagnética com a solução contendo o reagente cromogênico, sendo o sinal analítico a absorbância. 

Abraços!

Author

Bruna Soares Dionizio

Olá Célia!

Muito obrigada,

Acreditamos que a sensibilidade está relacionada ao fato que utilizando o Photometrix tem-se apenas o processamento da imagem digital adquirida, ou seja, está relacionada com a aparência da solução. Enquanto que no espectrofotômetro há interação da radiação eletromagnética com a solução contendo o reagente cromogênico, sendo o sinal analítico a absorbância. 

Abraços!

Author

Bruna Soares Dionizio

Olá Maria Célia.

Nós utilizamos uma caixa de madeira com um orifício e suporte para a câmara do celular. Dessa maneira, todas as imagens obtidas para amostras e padrões estavam sujeitas ao mesmo ambiente.  

Abraço.