Assessment of agricultural and forestry biomasses quality for bioenergy and biofuels applications by NIRS
Important efforts have been undertaken concerning the characterisation of biomass quality for bioenergy and/or biofuel conversion. Due to the limitations in availability and to avoid competition with food production. the biomass supply for bioenergy and the second generation biofuels will be essentially composed of multi-ressources. Multi-species NIRS calibrations are rare or almost inexistent in the specific literature. In this work, the prediction of biomass quality of agricultural and forest biomasses for bioenergetics applications by multi-ressource NIRS calibrations was carried out.
NIR spectra were acquired using a Bruker Tensor 37 equipped with an integration sphere and a rotating device allowing a homogenisation of the sample during spectra acquisition. Three spectra and 20 scans were acquired for each sample by in a spectral range from 4,000 to 10,000 cm-1 at a resolution of 8 cm-1. The principle of property prediction is based on a calibration correlating spectroscopic data to the laboratory measurements using PLS-based methods implemented in OPUS software furnished by Bruker with the NIR spectrometer. Principal Component Analysis (PCA) was also used to evaluate the distribution of the agricultural and forest samples in the calibrations.
Chemical composition (lignin, cellulose, hemicelluloses, hexoses and pentoses content), elemental analysis (C, H, O content), ash content and inorganic composition of the samples were predicted by NIRS. All the parameters will be presented and discussed in the conference. Some examples of NIRS performances are given here. Lignin content varied from 2 to 18 % for agricultural biomass and from 14 to 36 % for forestry biomass. Despite these differences, a very good calibration with R² value of 0.914 for the calibration and 0.892 for the RMSEP (root-mean-square-error of prediction), was 2.7 %, which is higher than the precision of the lignin content measurement in laboratory (1 %). Polysaccharides ranges from 35 to 80 %, but forest biomass are concentrated between 50 and 80 %. R² values of the NIRS calibration for the prediction of polysaccharides content are 0.914 for the calibration and 0.891 for the cross-validation. The RMSEP obtained was 3.2 % to be compared to 2 % of precision of the wet analysis in laboratory. Finally, an excellent NIRS calibration was produced for the ash content ranging from 0.6 to 11.5 %. For ash content higher than 6 %, only agricultural samples were present. R² values for calibration and cross-validation were respectively 0.963 and 0.886. RMSEP value was 0.8 %, comparable to the precision obtained in laboratory which is around 0.5 %.
The fact that RMSEP were higher than the precision obtained in laboratory is probably due to the fact that agricultural and forestry samples were pooled together in the same calibration. Also, different extraction methods prior lignin and polysaccharides content measurements were used according to the origin of the sample. Finally, the morphology of ground agricultural and forestry were also slightly different. These differences were evidenced by some segregation in the PCA carried out for the different quality criteria. Nevertheless, the feasibility of the NIR spectroscopy for the prediction of multi-ressources biomass was demonstrated.